ROCHE DIAGNOSTICS D-DIMER; FIBRIN SPLIT PRODUCTS Back to Search Results
Catalog Number 04912551190
Device Problem Low Test Results (2458)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/10/2016
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer received a questionable d-dimer result for one patient sample from a cobas c501 analyzer.The initial results were 0.12 and 0.13 ug/ml.The result from a bcs xp siemens system was 2806.06 ng/ml.The results were reported to the physician.The patient was not adversely affected.
 
Manufacturer Narrative
Sample from the patient was submitted for investigation.Based on investigation data, the customer's results from the cobas c501 were considered to be correct.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
D-DIMER
Type of Device
FIBRIN SPLIT PRODUCTS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key5463778
MDR Text Key39205473
Report Number1823260-2016-00229
Device Sequence Number1
Product Code GHH
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K062203
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 04/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2016
Device Catalogue Number04912551190
Device Lot Number61323101
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/12/2016
Initial Date FDA Received02/26/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/20/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age68 YR
-
-