Additional information received on 30mar2016 with the following: on (b)(6) 2016, the patient (age unknown) was positioned prone for a posterior cervical procedure.The patient was not repositioned at any time during the surgery.Sixty pounds of pressure was applied.The surgery was not performed with a stereotaxy device.Post surgery, the surgeon noticed the laceration on the patient's head on the single pin side of the skull clamp.The 2cm laceration required staples.Integra has completed their internal investigation on 29mar2016.The investigation included: methods: evaluation of actual device.Review of device history records.Review of complaint history.Results: evaluation of device: with respect to the returned unit it has passed all specific functional testing requirements, except for the lock having rotational movement.This would not have caused a slippage.When the unit is properly positioned and put under pressure, the unit would not have slipped.The unit needs to be machined to have heli-coils added to large starburst threads; general maintenance and cleaning required.Device history record reviewed for this product id work order and lot # 00593/139 manufactured on october 25, 2013 show no abnormalities related to the reported failure.This device passed all required inspection points with no associated mrr¿s, variances or rework.No service history is on file for this device.No manufacturing or design related trend has been identified.In summary, the end user's experience could not be confirmed or duplicated.The returned unit passed all functional testing requirements, however general maintenance is required as this device was manufactured in 2013 with no prior service history.
|