Model Number 840
Device Problem Poor Quality Image (1408)
Patient Problem No Patient Involvement (2645)
Event Date 02/01/2016
Event Type  malfunction  
Manufacturer Narrative
Covidien reference: (b)(4).
 
Event Description
It was reported that, during maintenance, the ventilator's lower display became dim.There was no patient involvement.
 
Manufacturer Narrative
(b)(4).The device was evaluated and the liquid crystal display (lcd) panel was replaced.
 
Manufacturer Narrative
The suspect component was returned to covidien/ medtronic¿s product analysis. a visual inspection of the returned component was performed.The returned component was installed into a test ventilator for analysis.An investigation was performed and the product analysis technician reported that the reported issue was verified.The identified root cause of the reported issue was an incorrect installation.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Type of Device
CONTINUOUS VENTILATOR
Manufacturer (Section D)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI 
Manufacturer (Section G)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI  
Manufacturer Contact
thom mcnamara
15 hamsphire street
mansfield, MA 02048
5084524811
MDR Report Key5465005
MDR Text Key39448584
Report Number8020893-2016-00450
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K970460
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number840
Device Catalogue Number840
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/01/2016
Initial Date FDA Received02/26/2016
Supplement Dates Manufacturer ReceivedNot provided
03/03/2016
Supplement Dates FDA Received04/08/2016
12/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/23/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-