The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The evaluation confirmed that the subject device did not turn on, two fuses were blown, and fet (field-effect transistor) of amplifier substrate was damaged and burnt.The manufacturing records for this serial number indicated no abnormalities related to the reported phenomenon.Based upon the evaluation result, it was supposed that this phenomenon occurred since the fet was damaged.It is considered that the fet was damaged since the electric current leaked excessively in the fet.Amount of the leaked electric current depends on the dispersion in the electrical characteristic of fet.Omsc addressed this phenomenon by reducing the dispersion of fet.The subject device instruction manual states the following.Prepare a spare instrument as a backup to ensure that the procedure can be completed without complications in case of a malfunction.
|
The subject device was used for a procedure of a heart blood vessel surgery.Though the user stepped on the foot switch, the activation didn't occurred several times.After that the user irrigated sonosurg-iu that was used together and stepped on the foot switch again.Then the subject device shut down with a strange sound, and smoke occurred.The procedure was completed with another device, and there was no patient injury reported.
|