• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES OBERDORF CHUCK WITH KEY FOR BATTERY POWER LINE; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES OBERDORF CHUCK WITH KEY FOR BATTERY POWER LINE; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.730
Device Problems Break (1069); Detachment Of Device Component (1104); Torn Material (3024)
Patient Problem No Patient Involvement (2645)
Event Date 01/20/2014
Event Type  malfunction  
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root cause was determined to be due to improper/faulty care and maintenance, which is user error, misuse, and / or abuse.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported from (b)(6) that during service and repair, it was observed that the gear was broken and torn off on the chuck device.This event was not related to surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHUCK WITH KEY FOR BATTERY POWER LINE
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ  CH4436
Manufacturer (Section G)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ   CH4436
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5467889
MDR Text Key39650714
Report Number8030965-2016-11064
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
Reporter Country CodeRS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 01/20/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.730
Device Lot Number7781007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2014
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/20/2014
Initial Date FDA Received02/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/29/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-