Model Number 840 |
Device Problem
Poor Quality Image (1408)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Covidien reference number: (b)(4).The service engineer (se) verified the malfunction.The se replaced the graphical user interface (gui) printed circuit board (pcb) and upgraded the backlight inverter pcb.The ventilator passed all testing.
|
|
Event Description
|
It was reported that the ventilator had a lower display with garbled and fragmented data.There was no patient connected to the device.
|
|
Manufacturer Narrative
|
The device was repaired and the reported issue was isolated to interface between the device and the failed component.If information is provided in the future, a supplemental report will be issued.
|
|