The manufacturer representative reported there was an impedance measurement of 1168 ohms between electrodes 2 and 3.The patient was implanted two months prior and their symptoms were returning.They experienced symptoms of nausea and vomiting.The physician calculated a new voltage with 1168 ohms and 5 ma.However they could not program a new voltage with 7.5 ma.There was question of whether this was related to the high impedance.Regarding the cause of the high impedance, it was inquired if it was possibly due to inflammation around the lead, as they were recently implanted and if it could be temporary.In addition, it was wondered if a lead break should be considered even if the impedance was not greater than 4,000 ohms.No cause was determined.An x-ray was advised to check the position of the leads.In addition, further actions included increasing the current to 10 ma and watching the impedance measurement.No further information was provided.A follow up report will be sent if additional information is received.
|