• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASTORA WOMEN'S HEALTH LLC AMS MONARC SLING SYSTEM; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASTORA WOMEN'S HEALTH LLC AMS MONARC SLING SYSTEM; MESH, SURGICAL, POLYMERIC Back to Search Results
Catalog Number 72403831
Device Problems Stretched (1601); Separation Failure (2547)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/25/2016
Event Type  malfunction  
Event Description
It was reported that during surgery, the monarc sling sheaths would not come off after implantation without extreme difficulty.After the monarc was placed, the sling was cut below the blue dot, and the mesh sleeve was tried to be removed with a haemostat but would not come off.The physician then tried the other side and experienced the same difficulty.The tensioning suture then broke and the mesh became stretched while trying to remove the plastic sheaths.The physician decided to remove the whole device and used another monarc with no problem.No further complications were noted.If additional information is received, a follow up report will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS MONARC SLING SYSTEM
Type of Device
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
ASTORA WOMEN'S HEALTH LLC
13200 pioneer trail
suite 100
eden prairie MN 55347
Manufacturer (Section G)
ASTORA WOMEN'S HEALTH (IRELAND)
athlone business & tech park
garrycastle, dublin road
co. westmeath
EI  
Manufacturer Contact
erika a. merrick
13200 pioneer trail
suite 100
eden prairie, MN 55347
9522383906
MDR Report Key5470413
MDR Text Key39417842
Report Number3011770902-2016-00094
Device Sequence Number1
Product Code OTN
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 02/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/30/2018
Device Catalogue Number72403831
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/25/2016
Initial Date FDA Received03/01/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/24/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-