DATEX-OHMEDA, INC. AISYS Back to Search Results
Model Number AISYS
Device Problems Loss of Power (1475); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem Therapeutic Effects, Unexpected (2099)
Event Date 11/10/2015
Event Type  Injury  
Event Description
The vaporizer for the anesthesia machine was switched out during surgery.The alarm on the machine and the vaporizer shut off automatically after a short period of time.The process is reliant on human factors, i.E.The anesthesiologist to manually acknowledge and activate the vaporizer.Due to a distraction, anesthesia did not notice the vaporizer had shut off and there was no additional warning from the machine.Patient was not getting the agent necessary to maintain anesthesia and woke during the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS
Type of Device
AISYS
Manufacturer (Section D)
DATEX-OHMEDA, INC.
pittsburg PA 15264
MDR Report Key5471179
MDR Text Key39440678
Report NumberMW5060644
Device Sequence Number1
Product Code BSZ
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Nurse
Type of Report Initial
Report Date 02/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAISYS
Other Device ID NumberKNO34197
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/26/2016
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age17 YR
Patient Weight102
-
-