Follow-up #1: date of submission 03/17/2017 device evaluation: the device has been returned and evaluated by product analysis on 02/24/2017 with the following findings: a review of the black box showed the last basal and bolus deliveries, and a replace cartridge alarm with deliveries resumed shortly after.The total daily doses added up correctly and reflected the user's programmed basal rates.The pump passed delivery accuracy testing and was found to be delivering accurately and within range.The rewind, load, and prime steps were performed successfully in addition to the 24 hour testing.No delivery interruptions or alarms occurred during the investigation.The complaint of inaccurate delivery was unable to be duplicated.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.(b)(4).
|