Patient information was not made available from the site.On (b)(4) 2016 - a medtronic representative performed a navigation system check-out, all areas passed.System performed as intended.Return requested for suspect tap 4.5.Medtronic investigation of returned suspect device finds that, as reported, the tip of the instrument is slightly bent.The tip is also blocked with bony material.Physical damage - bent instrument tip hardware investigation was completed.This issue was found related to a hardware issue and was documented in a medtronic hardware anomaly tracking database.No parts have been received by manufacturer for analysis.No further issues have been reported.
|
A medtronic representative received a report from a site that, while in a spine procedure, they discovered a 4.5 solera tap that was visibly bent near the tip.The instrument passed verification, however, when the tap was rotated by the site surgical technician they noticed it was visibly bent.This damage was noted prior to the tap being used in the procedure.The surgeon opted to continue and completed the procedure with the use of the navigation system.Delay in therapy was less than one hour.There was no impact on patient outcome.
|