Catalog Number 273205110
Device Problem Torn Material (3024)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/11/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4) a follow up report will be filed upon completion of the investigation.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Rep reported this complaint to synthes.Reporting a couple damaged instruments and trial that occurred in a case today, (b)(6).Cougar inserter: both cross threaded in case 273511300 lot: ag2066621.273511300 lot: ag2066617.Cougar 10mm 5deg small trail: medial lateral trial threads cross threaded 273205110 lot: t0911, it was reported that during an anterior lumbar interbody fusion (alif) at l4-l5 on (b)(6) 2016, the inserter tip stripped.When inserting the second inserter into the trial, the inserter tip stripped and the trial became damaged.No fragments generated.Another set was available and surgery was successfully completed without patient harm.A four minutes surgical delay was reported.Devices and implant are available for return.Sales consultant confirmed no additional information.
 
Manufacturer Narrative
One (1) small 5 degree trial (product code: 2732-05-110, lot no: t0911) was returned to the complaint handling unit (chu) for evaluation.Visual examination of the returned trial revealed notable signs of wear and the leading thread on two of the trial hole threads are significantly worn-out.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.No emerging trends were found requiring further actions.With the information provided, a definitive root cause for torn threads on the cage inserter and the small 5 degree trial cannot be determined.However, possible root cause of torn threads may be due to incorrect component positioning or inadequate thread engagement of the inserter into the small trial and inadvertently cross threading.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends requiring immediate action have been observed, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAGE TRIAL, SMALL
Type of Device
TEMPLATE
Manufacturer (Section D)
DEPUY SYNTHES SPINE
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY SYNTHES SPINE
325 paramount drive
raynham MA 02767
Manufacturer Contact
michael jacene
325 paramount drive
raynham, MA 02767
5089776485
MDR Report Key5484405
MDR Text Key40036700
Report Number1526439-2016-10185
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number273205110
Device Lot NumberT0911
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location Hospital
Initial Date Manufacturer Received 02/11/2016
Initial Date FDA Received03/07/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/06/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/04/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-