• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET ORTHOPEDICS G7 10 DEG E1 LINER 40MM I; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET ORTHOPEDICS G7 10 DEG E1 LINER 40MM I; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Component or Accessory Incompatibility (2897)
Patient Problem No Information (3190)
Event Date 02/11/2016
Event Type  Injury  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.There are warnings in the package insert that state that this type of event can occur.This report is number 3 of 3 mdrs filed for the same event (reference 1825034-2016-00818 / 00820).
 
Event Description
It was reported the patient underwent a total hip arthroplasty on an unknown date in 1992.Subsequently, the patient was revised on (b)(6) 2016 due to wear of the polyethylene liner.All components were removed and replaced.During the procedure, two liners would not seat in the cup, resulting in a 45 minute delay in procedure.
 
Manufacturer Narrative
This follow-up report is being filed to relay additional information, which was unknown at the time of the initial medwatch.Examination of returned device found no evidence of product non-conformance.Review of the device confirmed the reported condition.Deformation of the device was noted.A conclusive root cause of the event could not be determined.Corrective action has been initiated to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 10 DEG E1 LINER 40MM I
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
megan haas
56 e. bell drive
warsaw, IN 46582
5743726700
MDR Report Key5487048
MDR Text Key39927739
Report Number0001825034-2016-00820
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date04/30/2020
Device Model NumberN/A
Device Catalogue Number010000906
Device Lot Number3559273
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/11/2016
Initial Date FDA Received03/08/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/21/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/29/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-