It was reported that, patient underwent posterior lumbar inter-body fusion surgery for isthmic spondylolisthesis at l4-l5.During the surgery, it was found through image that the right screw at l5 was placed "too medially" after insertion of screw so the screw was removed and inserted again.Image found no problem after the re-insertion.Post-op, it was confirmed non smooth movement of legs so the surgical site was checked by ct.It was observed the re-inserted screw has come out on the inside and revision surgery was performed on the same day.The surgeon commented that confirmation in the x-ray was difficult due to patient's bone quality.Surgeon also commented that he should have considered the need to use navigation too in such situation.Product came in contact with the patient.No patient complications were reported.
|