• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD; SWAN-GANZ CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD; SWAN-GANZ CATHETER Back to Search Results
Model Number 774HF75J
Device Problems Knotted (1340); Physical Resistance (2578)
Patient Problem No Code Available (3191)
Event Date 02/12/2016
Event Type  Injury  
Manufacturer Narrative
No product was returned for evaluation; it was discarded at the hospital.Without the return of the product, it is not possible to determine if damages or defects existed on the product.The lot number was not provided; therefore, a review of the manufacturing records could not be completed.No actions will be taken at this time.
 
Event Description
It was reported that while attempting to remove the catheter, resistance was felt.A cut-down was performed and the catheter was removed with the introducer.It was noted that a knot was formed at the tip of the catheter.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD
Type of Device
SWAN-GANZ CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn selawski
1 edwards way
irvine, CA 92614
9497564386
MDR Report Key5490371
MDR Text Key40027556
Report Number2015691-2016-00728
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 02/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number774HF75J
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/12/2016
Initial Date FDA Received03/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-