INTERSECT ENT PROPEL SINUS IMPLANT; DRUG ELUTING SINUS STENT Back to Search Results
Model Number 70011
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Choking (2464)
Event Type  malfunction  
Manufacturer Narrative
The physician stated the implant was placed correctly, however the patient's anatomy may have contributed to the migration.In an abundance of caution, intersect ent is reporting this event as a malfunction based on the information received and the remote possibility of a serious injury should a similar event occur.(b)(4).
 
Event Description
The patient underwent revision endoscopic sinus surgery (ess) in (b)(6) 2016.Sinus implants were placed bilaterally, in previously modified anatomy, in the ethmoid sinuses without incident.Approximately one week post-surgery the implant on the right side had migrated to the patient's throat area; the left side implant remained in place.The patient felt something in their throat (described as a choking sensation) and concerned about a possible obstruction of the airway went to the er.The reporting physician stated that the patient coughed out the implant prior to receiving intervention while in the er; patient is fine.
 
Manufacturer Narrative
Voluntary event report (mw 5060385) which contained additional information was received on 3/24/2016.Additional narrative: the information provided in the initial report indicated that the patient visited er.Clinical quality, safety & risk specialist from the hospital confirmed the patient did not mention an er visit when reporting the event to them.Additionally, the hospital had no record of the patient receiving treatment in their er.Based on the lhr review, the product met all specifications.A potential root cause related to manufacturing / process control could not be identified.
 
Event Description
The patient underwent revision endoscopic sinus surgery (ess) (b)(6) 2015.Sinus implants were placed bilaterally, in previously modified anatomy, in the ethmoid sinuses without incident.Approximately one week post-surgery while the patient was driving, the implant on the right side had migrated to the patient's throat area; the left side implant remained in place.The patient reported to the hospital that he pulled off the freeway and removed the object from his throat manually and then proceeded to drive on to their post-operative appointment with the physician.Patient is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROPEL SINUS IMPLANT
Type of Device
DRUG ELUTING SINUS STENT
Manufacturer (Section D)
INTERSECT ENT
menlo park CA
Manufacturer (Section G)
INTERSECT ENT
1555 adams dr
menlo park CA 94025
Manufacturer Contact
amy wolbeck
1555 adams dr
menlo park, CA 94025
6506412115
MDR Report Key5490510
MDR Text Key40031806
Report Number3010101669-2016-00003
Device Sequence Number1
Product Code OWO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100044
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date08/28/2017
Device Model Number70011
Device Catalogue Number70011
Device Lot Number50828001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/18/2016
Initial Date FDA Received03/09/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/21/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/28/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-