MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS, INC MERIT CUSTOM KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS, INC MERIT CUSTOM KIT Back to Search Results
Catalog Number K04-MZL3501
Device Problem Unsealed Device Packaging (1444)
Patient Problems No Patient Involvement (2645); Not Applicable (3189)
Event Date 11/19/2015
Event Type  malfunction  
Manufacturer Narrative
One device was returned for evaluation.The evaluation is in process.A follow up report will be submitted when the evaluation has been completed.
 
Event Description
The distributor reported a defect in the packaging.This was identified during their initial inspection of received product.The device was not sent to a user facility.
 
Manufacturer Narrative
One device was returned for evaluation.On visual inspection it was observed that one of the items in the kit was caught in the packaging seal leading to a breach in the sterile packaging.The complaint is confirmed.The root cause is attributed to incorrect placement of components inside the packaging prior to sealing the package.The device history record was reviewed and no exception documents were found.The complaint database was reviewed and no similar complaints for this lot number were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERIT CUSTOM KIT
Type of Device
CUSTOM KIT
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC
1600 west merit parkway
south jordan UT 84095
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS, INC
1600 west merit parkway
south jordan UT 84095
Manufacturer Contact
jerry mcphie
1600 west merit parkway
south jordan, UT 84095
8012084491
MDR Report Key5491166
MDR Text Key40024681
Report Number1721504-2016-00033
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 02/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2018
Device Catalogue NumberK04-MZL3501
Device Lot NumberH833460
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/09/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/16/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-