U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC. LAP-BAND AP STANDARD W/RAPIDPORT EZ; ADJUSTABLE GASTRIC BAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC. LAP-BAND AP STANDARD W/RAPIDPORT EZ; ADJUSTABLE GASTRIC BAND Back to Search Results
Model Number N/A
Device Problems Fluid/Blood Leak (1250); Inadequacy of Device Shape and/or Size (1583)
Patient Problem Patient Problem/Medical Problem (2688)
Event Type  malfunction  
Manufacturer Narrative
Strain relief (rapidport ez).Medwatch sent to fda on 03/10/2016.The device was returned to apollo for analysis.A visual analysis was performed on the returned lap-band and noted the device was discolored, and it appeared blue in color.Particles were noted on the inner surface of the band.An air leak test was performed and noted leakage at the junction of the buckle and shell.A microscopic analysis was then performed on the device, and noted an unidentified opening between the shell and the buckle.Device labeling addresses possible outcomes of band restriction issues and leak(s) as follows: adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Unplanned deflation of the band may occur due to leakage from the band, the port or the connecting tubing.
 
Event Description
Reported as: a patient with the lap-band system was reported to have come "in due to no restriction.The reporter was unsure if any diagnostics tests were done prior to surgery.Surgery was done to evaluate if patient needed a revision.It was discovered that the lap-band was leaking (a hole) either between the buckle and the ring or the belt and the inner shell.The lap-band system was replaced.
 
Manufacturer Narrative
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND AP STANDARD W/RAPIDPORT EZ
Type of Device
ADJUSTABLE GASTRIC BAND
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s. captial of texas hwy
bldg 1, ste. 300
austin TX 78746
Manufacturer (Section G)
ALLERGAN
global park free zone
900 global park
la aurora de heredia, costa rica, cs
CS  
Manufacturer Contact
laura leboeuf
1120 s. capital of texas hwy
bldg 1, ste 300
austin, TX 78746
5122795141
MDR Report Key5493698
MDR Text Key40121358
Report Number3006722112-2016-00061
Device Sequence Number1
Product Code LTI
UDI-Device Identifier10811955020268
UDI-Public10811955020268
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date12/31/2016
Device Model NumberN/A
Device Catalogue NumberB-2360
Device Lot Number2688174
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/11/2016
Initial Date FDA Received03/10/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/29/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/21/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age54 YR
-
-