The reported incident that 10 hole pin clamp hoffmann ii mri 4-6mm was alleged of issue s-31 (no fitting) could be confirmed, since the returned device matched the reported failure.Based on investigation, the root cause was attributed to a user related issue.The failure was probably caused by mishandling of the device.The device inspection revealed that the sample post cannot be inserted in one of the holes in the green side of the pin clamp, confirming the reported event.This hole is deformed inside, which points to the multiple usage of this device.These deformations are what is stopping the post to be fully inserted in the pin clamp.Multiple use devices need to be carefully inspected prior and post surgery, since damages can have occurred in the last usage and so screening is needed not to reuse damaged parts.Also note, that this clamp should not be tightened without the post inserted, otherwise deformation can occur that can compromise the performance of the device.Note, as stated in the operative technique (lh2mri-ot rev 1 hoffmann ii mri optech): ''note: do not over torque the bolt b which does not contain a post.'' [original statement(s)].Note, as stated in the ifu (v15034): ''ensure that you are familiar with the intended uses, indications/contraindications, compatibility and correct handling of the external fixation devices, which are described in the operative technique manual for the product system.'' [original statement(s)].A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
|