Model Number 16414
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 02/17/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation is in progress, but not yet concluded.The fsr opened unit and found connector was completely detached from the board.The customer was not currently using this pressure, but perfusion wanted it repaired.The fsr installed a new cpg temp/press board, reassembled the monitor and completed testing.The unit operated to manufacturer specifications and was returned to clinical use.The suspect part was returned to the manufacturer for further evaluation.
 
Event Description
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, he found the "b" pressure on the cardioplegia (cpg) monitor would not zero and the connector's orientation was not correct.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 CARDIOPLEGIA MONITOR
Type of Device
COMPUTER, BLOOD-PRESSURE-SARNS 8000 CARDIOPLEGIA MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5494672
MDR Text Key40405201
Report Number1828100-2016-00173
Device Sequence Number1
Product Code DSK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960916
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number16414
Device Catalogue Number16414
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/17/2016
Initial Date FDA Received03/11/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/25/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-