During a generator replacement for battery depletion, high impedance was observed when the existing lead was connected to the new generator.Interrogation of the existing generator prior to surgery was unsuccessful despite multiple attempts due to the generator battery being depleted.During the surgery, the surgeon visible saw a lead fracture below the anchor tether and also noted fluid in the lead.Therefore the lead was also replaced.Diagnostics were within normal limits after the revision surgery (dcdc-1).The explant facility do not return the explanted products back to the manufacturer.
|