Catalog Number J2C1711
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Patient Involvement (2645)
Event Date 02/15/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4) - the device was manufactured august 25, 2015 ¿ august 26, 2015.The device was received for evaluation.Visual inspection revealed that the stressmemeber flange was loose.The cause of the condition was determined to be an assembly issue during manufacturing.During the assembly step, the groove of the stressmember flange must be aligned correctly onto the groove of the housing in order for the stressmember flange not to move or become loose.A capa has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a small volume infusor had a loose stress member.This was observed before use.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
bldg 3
irvine CA 92614
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5496174
MDR Text Key40202483
Report Number1416980-2016-05041
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/11/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date08/01/2018
Device Catalogue NumberJ2C1711
Device Lot Number15H040
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/17/2016
Initial Date FDA Received03/11/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-