(b)(4) - the device was manufactured august 25, 2015 ¿ august 26, 2015.The device was received for evaluation.Visual inspection revealed that the stressmemeber flange was loose.The cause of the condition was determined to be an assembly issue during manufacturing.During the assembly step, the groove of the stressmember flange must be aligned correctly onto the groove of the housing in order for the stressmember flange not to move or become loose.A capa has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|