Submit date: 03/10/2016.A review of the information in the complaint file indicates this investigation was performed by a medtronic technical center for the reported condition of; the unit is not working.During triage the unit was found to have a damaged power cord that was cut, burnt, and had exposed copper wire.Therefore, this report will be based on information provided by the technical center.The scd express was evaluated and the customer reported issue was confirmed; the unit's main board had solder shorts at u4 and the power cord's outer insulation was damaged, allowing view of the inner copper wire.The cause of the reported condition for the damaged power cord and the unit not working was due to customer misuse.The unit's main board and power cord were scrapped and replaced to correct the reported issue.The unit passed all initial testing after failure analysis repair instructions were completed.Scd express was manufactured in 2007.A review of the device history record shows this device was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.Based on the investigation, no corrective action is needed.
|