• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INC. NUVASIVE PRECEPT SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INC. NUVASIVE PRECEPT SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 8800000006
Device Problem Detachment Of Device Component (1104)
Patient Problem Failure of Implant (1924)
Event Date 02/18/2016
Event Type  malfunction  
Manufacturer Narrative
Radiographs received confirmed the separation of three lock screws.The patient underwent revision surgery with no reported complications.No further investigation can be completed at this time.Root cause has not been determined, no conclusion can be drawn.Review of labeling notes: potential adverse events and complications: "potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s)and loss of fixation." warnings, cautions and precautions: "confirm heights of screws match with patient's lordosis when securing lock screws.Failure to verify screw height following insertion may result in inadequate rod normalization." "final-tighten all lock screws with the counter-torque and torque t-handle.Do not final-tighten through other instruments in the set, as the rod may not be able to normalize to the tulip.Ensure there is enough rod overhang when final tightening, and do not lock down on the conical portion of the rod.Failure to do so may lead to improper lock down of the construct.It may be necessary for the surgeon to add length to the rod, depending upon patient anatomy and desired lordosis." not returned.
 
Event Description
In (b)(6) 2016, a female patient underwent posterior fixation at l3-l4 to treat vertebral osteomyelitis.On (b)(6) 2016 surgeon reported that radiographs showed three set screws had disassociated from the construct.Revision surgery was performed on (b)(6) 2016 and the fixation was extended to s2.Patient is reported as doing well after revision surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE PRECEPT SPINAL SYSTEM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
NUVASIVE, INC.
7475 lusk blvd
san diego CA 92121
Manufacturer Contact
stacy gregory
7475 lusk blvd
san diego, CA 92121
8587360313
MDR Report Key5498372
MDR Text Key40252374
Report Number2031966-2016-00019
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122352
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model Number8800000006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/25/2016
Initial Date FDA Received03/14/2016
Was Device Evaluated by Manufacturer? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
-
-