• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; HEATED HUMIDIFICATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; HEATED HUMIDIFICATION SYSTEM Back to Search Results
Catalog Number 425-00
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 03/04/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the device was not returned at the time of this report.The device history record (dhr) review did not show issues related to the complaint.The device was manufactured on 09/23/2008.A document assessment ((b)(4)) was conducted and no changes were required.The reported complaint cannot be confirmed since the device sample is not available to perform a proper investigation, determine the root cause and assign corrective actions.If the device sample becomes available at a later date, this complaint will be updated accordingly.
 
Event Description
The event is reported as: the panel display flickers prior to use.There was no patient involvement reported.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and there were no defects observed.Functional testing was performed and the unit was connected to 110vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.Based on the investigation performed, the reported complaint of "panel display flickers" could not be confirmed.Functional testing did not reveal any operational anomalies.
 
Event Description
The event is reported as: the panel display flickers prior to use.There was no patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
HEATED HUMIDIFICATION SYSTEM
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key5504928
MDR Text Key40496516
Report Number3003898360-2016-00290
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number425-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/04/2016
Initial Date FDA Received03/16/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/07/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-