Model Number MODEL 250
Device Problems Sticking (1597); Failure to Advance (2524)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/04/2016
Event Type  malfunction  
Event Description
On (b)(6) 2016 it was reported that the handheld is stuck on the alignment screen.The nurse has tapped the crosshairs over 20 times and it is not advancing.It was reported that the handheld would be returned for product analysis but it has not been received to date.
 
Event Description
On 4/20/2016 the handheld and flashcard were returned for product analysis.The flashcard and software performed according to functional specifications.Product analysis on the handheld revealed that the handheld was unable to advance past the screen alignment utility.The cause for the display anomaly is associated with resistance values being higher than expected in the touch screen circuitry.Following replacement with a known good screen, full product functionality was restored.No further anomalies associated with the handheld performance were identified during the analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5513544
MDR Text Key41144619
Report Number1644487-2016-00601
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 03/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 03/04/2016
Initial Date FDA Received03/21/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/16/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/06/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-