11/8/16 integra investigation completed.Method: failure analysis, device history evaluation.Results: failure analysis - forceps returned in used condition, not showing any unusual markings.The returned forceps showing wear, staining, and a broken jaw.During the evaluation of the instrument, it is noticed that there is rust formation where the breakage appears to be.It appears the breakage may have started out as a fracture leading to the breakage.It is also noticed that the jaw is broken., the cause is undetermined.The complaint is confirmed.Device history evaluation - dhr review was completed with all history available.Nonconforming product report / nonconforming material report history: there is no applicable nonconforming product report/nonconforming material report history.Variance authorization / deviation history: none.Engineering change order/manufacturing change order history: there is no applicable engineering change order/manufacturing change order history.Corrective action preventive action history: none.Health hazard evaluation history: none.Conclusion: the root cause has not been identified as a workmanship or material deficiency.
|