• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CUSTOM ULTRASONICS INC. SYSTEM 83 PLUS 2; ENDOSCOPE WASHER - DISINFECTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CUSTOM ULTRASONICS INC. SYSTEM 83 PLUS 2; ENDOSCOPE WASHER - DISINFECTOR Back to Search Results
Device Problem Device Inoperable (1663)
Patient Problem No Information (3190)
Event Date 03/03/2016
Event Type  malfunction  
Manufacturer Narrative
There are no patient injuries associated with this event.The reported symptom for the returned ultrasonic generator (500 w) board was attributed to a faulty diode, on the ac input signal, within the bridge rectifier, brg1, on the ultrasonic generator (500 w) board, p/n 12008, causing the fuse to open, rendering the board inoperable.This diode is located on the ac input plus side of the bridge rectifier and its failure is likely attributed to age (16 years) and an over current condition.The loss of the ultrasonic generator does not, of and by itself, result in an endoscopic high level disinfection process failure that does not render a microbiologically safe product or present any additional risks for reuse given the nature of the pre-cleaning labeling instructions and high-level disinfection.Given the nature of endoscope pre-processing and high-level disinfection based on sgna guidelines, either the presence or absence of ultrasonics should still result is an endoscope that has had its bioburden reduced to an acceptable level.It is concluded that there is no reasonable likelihood that an endoscope processed in the absence of the ultrasound will present any additional risks.
 
Event Description
During a scheduled preventive maintenance visit the technician noticed the ultrasonic function was not working.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYSTEM 83 PLUS 2
Type of Device
ENDOSCOPE WASHER - DISINFECTOR
Manufacturer (Section D)
CUSTOM ULTRASONICS INC.
144 railroad drive
ivyland PA 18974
Manufacturer Contact
elizabeth lazaro
144 railroad drive
ivyland, PA 18974
2153641477
MDR Report Key5520141
MDR Text Key41256684
Report Number3007082252-2016-00005
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K983017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/03/2016
Initial Date FDA Received03/23/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/19/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-