(b)(4).The handpiece was received attached to a harh36 shaft.The handpiece was forcibly removed from the disposable and the handpiece got disassembled.Upon visual inspection it was observed that the nose cone was cracked and the acoustic isolator was torn.Due to the condition of the device, no functional test could be performed.The transducer assembly was not held in place due to the cracked nose cone, so torquing on the disposable resulted in twisting only the handpiece transducer assembly until the internal wires got disconnected.The end cap was disassembled to inspect remaining components and it was found that the moisture indicator was positive.In addition, degradation of the hand activation wire outer jacket was observed.Due to the cracked nose cone, moisture entered the hand piece mid housing.¿positive moisture indicator¿ describes a condition where water enters the handpiece cavity during the steam sterilization process due to the damaged nose cone.A possible cause of the nose cone being cracked is the sterilization method due to the heating and cooling of the sterilization cycles is a stressor.The batch record was reviewed and no anomalies were noted during the manufacturing process.
|
It has been reported that during an unknown procedure, the handpiece had been used with the disposable device and was not working.Indeed, after screwing the two pieces together it was not activating and no error message appeared.Finally the two devices can not be taken apart.
|