(b)(4).There are several potential patient and procedural factors that alone or in combination can cause or contribute to a report of a restricted or nonfunctioning leaflet.Based on historical review of complaints, these events are typically a result of too ventricular deployment of the valve in combination with native leaflet overhang.Other potential contributing factors include: leaflet impingement in a highly calcified native valve, impingement of a leaflet due to the guide wire, or slow recovery of adequate ventricular flow post valve deployment and rapid pacing.This can result in a temporary decrease in the pressure gradient between the ventricle and the aorta, resulting in an inadequate pressure change to close the leaflets.In many instances this can be overcome with trouble shooting, which includes blood pressure recovery or support.Occasionally there are cases where the root cause of the nonfunctioning leaflet cannot be determined.Per the instructions for use, central regurgitation is a potential adverse event associated with bioprosthetic heart valves and the transcatheter aortic valve replacement (tavr) procedure.There are multiple patient and procedural factors that alone or in combination can cause or contribute to central regurgitation including malposition of the valve, impingement of a leaflet due to the guide wire, over inflation of the deployment balloon, post dilation of the implanted valve, and slow recovery of adequate ventricular flow post valve deployment and rapid pacing.All of these factors have the potential to contribute to suboptimal coaptation of the sapien valve leaflets and cause central aortic insufficiency.Occasionally there are cases where the root cause of the regurgitation cannot be determined.The thv training manuals instruct the operator on proper positioning and deployment of the valve, including all procedural and anatomical considerations.The patient screening manual instructs the operator on proper native valve leaflet assessment, taking into consideration the length, bulkiness and distribution of calcium on the native leaflets to determine whether valve performance will be impaired.During the manufacturing process, all sapien valves are 100% visually inspected for defects and 100% tested for coaptation prior to release for distribution.This makes it highly unlikely that a manufacturing defect or device malfunction would contribute to the event.In this case, the exact cause of the severe central ai and motion restricted leaflet could not be confirmed, but maybe related to the mechanisms described above.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.
|
During the transfemoral tavr procedure, post deployment of a 23mm sapien 3 valve in a 60:40 aortic position, echo showed significant central ai and a leaflet that was ¿not working properly¿.The team advanced a pigtail catheter across the valve and slowly moved it in and out.A repeat echo showed slight improvement.A 23x4 balloon was then used for post dilated.Echo findings confirmed the central ai as mild-moderate.The decision was made to deploy a second valve.A 23mm sapien 3 valve was deployed within the first valve, which reduced the central ai to none.The patient remained stable throughout the procedure.As of pod9 the patient remains in stable condition.
|