CODMAN & SHURTLEFF, INC. CODMAN DISPOSABLE PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED) Back to Search Results
Model Number 26-1221
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/25/2015
Event Type  malfunction  
Event Description
During left temporal craniotomy for temporal mass suspected brain tumor, the surgeon was creating 4 burr holes.On the last burr hole creation the perforator did not stop.The surgeon immediately pulled the drill back without injury to the underlying dura.The perforator has built in stop mechanism so that when the inner portion of the skull is breached it is supposed to stop.The drill's diameter is approximately the size of a nickel.This failure could have injured the brain matter.There was no delay in the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOSABLE PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
325 paramount drive
raynham MA 02767
MDR Report Key5531186
MDR Text Key41313670
Report Number5531186
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 03/23/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date03/31/2020
Device Model Number26-1221
Device Lot NumberDH004S
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/23/2016
Device Age0 DY
Event Location Other
Date Report to Manufacturer03/23/2016
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/29/2016
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age50 YR
-
-