MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP-JAPAN
Device Problem Air Leak (1008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/05/2016
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that there was an air leak coming from the tip of the cutter during an eye surgery.The product was replaced and the procedure was completed without consequences to the patient.A product sample has been requested for evaluation.
 
Manufacturer Narrative
A product sample was returned for evaluation.The final lot was identified.A device history record review for the lot was conducted.The product was released based on the product¿s acceptance criteria.The sample was visually inspected using 25x magnification and deemed conforming.Actuation and aspiration testing were performed and deemed conforming.No air leakage was observed out of the port.The complaint evaluation does not confirm air leakage from the tip port.The probe was manufactured and functioned to specification.The root causes for the complaint issues are unknown because the returned sample was conforming to specification.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
eddie darton, md, jd
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175686660
MDR Report Key5531940
MDR Text Key41652395
Report Number2028159-2016-01347
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP-JAPAN
Device Catalogue Number8065751817
Other Device ID Number3.00.61
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/06/2016
Initial Date FDA Received03/29/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/22/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-