It was reported that the central venous pressure (cvp) was not correct according to the patient status.During set up of the device, the flushing and the system worked correctly and the values were correct (10 mmhg) when the monitoring started in the operating room.However, after 30 minutes the pressure decreased to minus 17 mmhg.There was no alarm.Furthermore, all external possibilities that could have influenced the pressure were checked, as placing of the sensor level, filling and pressure (300 mmhg) of the saline, air bubbles, kinking of the line, stopcocks open/closed, but no cause was found.Replacing the device for another one solved the issue; thus the values displayed came back to 10-11mmhg.There was no allegation of patient injury.The product is available for evaluation.
|