The batch documents were reviewed and the records showed no visual and functional problem on the batch during the manufacturing processes.The review confirmed that the operation, materials and tests were performed per the manufacturing processes.All transducers gone through electrical test and there was no indication of similar issue occurred during the manufacturing processes.The product passed and the batch was accepted and released.Further analysis was conducted on the cable connectors.On one (1) transducer cable it was observed that two pins which were very close to each other were intermittently contacting each other.The insulation heat shrink tube on one of the wire was too short and hence, not able to fully insulate the exposed wire.There is no visual and functional problem identified on the other transducer unit.Feedback and awareness from cable supplier was initiated for the effectiveness improvement action at cable connection jacket.
|
The kit filled as prescribed reset to 0 and will then be in use.The patient has high blood pressure rises , and despite treatment continues to increase.In another reset of the pressure line , it is not possible to reset to 0 , and therefore be taken down.A new set is used and then seen as low blood pressure that may be given medication to get an acceptable blood pressure.
|