During a prophylactic generator replacement (ifi = yes) on (b)(6) 2016, the surgeon accidentally cut the lead and had to perform lead revision.The explanted generator and lead were received.An analysis was performed on the returned lead portions.The reported allegations of lead fracture due to surgeon accidentally cutting the lead was confirmed based on the condition of the returned lead portions.During the visual analysis of the returned lead, abraded openings were observed on the connector pin and connector ring inner silicone tubes.Pitting was observed on the connector ring quadfilar coil surface, in this area.Analysis confirmed opening of both inner tubing sections in adjacent areas, exposing conductive quadfilar coils and resulting in coil pitting and a potential short-circuit condition.With the exception of the slice marks, puncture marks and abraded inner tubing openings, the condition of the returned lead portions is consistent with those that typically exist following an explant procedure.The abraded openings of both inner tubes is suspected to have occurred during the implant life and prior to explant surgery.The setscrew marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portions were performed, during the visual analysis, and no discontinuities were identified.Based on the findings, there is no evidence to suggest discontinuities in the returned portions of the device which may have contributed to the stated allegations of high impedance/lead fractures.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.The battery, 2.914 volts, shows an ifi=no condition.Various electrical loads were attached to the pulse generator and results of diagnostic tests demonstrate that accurate resistance measurements were obtained in all instances.There were no performance or any other type of adverse conditions found with the pulse generator.
|