• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number MERGE HEMODYNAMICS 9.40
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Code Available (3191)
Event Date 03/03/2016
Event Type  malfunction  
Manufacturer Narrative
Based upon the available information, there is no indication that the reported problem was due to merge healthcare's device malfunctioning.The customer performed simulation testing and found that the problem was due to a cabling issue with an external system.(b)(4).
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that problems were experienced with the ekg output to a volcano device.Information obtained from the customer revealed that the problem occurred while a transplant patient was sedated, and the procedure was delayed 20 minutes.It was also reported that the heart catheterization procedure was completed successfully; however, a flow wire study was attempted but cancelled due to the problem.The flow wire study will be completed at the patient's next visit - usually an annual visit unless problems present before then.It was further reported that the problem was not caused by merge hemo but rather a cabling issue with an external system.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge drive
hartland, WI 53029
2629123570
MDR Report Key5543003
MDR Text Key41719924
Report Number2183926-2016-00522
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/03/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS 9.40
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/03/2016
Initial Date FDA Received04/01/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/21/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-