Other # field is populated with the di number.Additional suspect medical device components involved in the event: model #: sc-2138-50, serial #: (b)(4), description: scs 50cm iii, lead sc-1110, sn (b)(4): device evaluation indicated that the ipg passed all tests performed.The ipg exhibited normal device characteristics.Sc-2138-50, sn (b)(4).Device evaluation indicated that the lead passed all tests performed.The device exhibited normal device characteristics.Sc-2138-50, sn (b)(4).Device evaluation indicated that the visual (microscope) and x-ray inspection of the lead revealed that multiple of the cables were completely broken at the bent/kinked location of the lead.This location appeared to be the site where the suture was positioned.No cables were exposed at the fracture site.
|