MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (COYOL) OPTICROSS¿; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (COYOL) OPTICROSS¿; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number H749518080
Device Problem Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/19/2016
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: the device was returned for evaluation.Device analysis revealed an open hole in the lapjoint area.A kink was observed in the sheath assembly from femoral marker to the distal end.Fluid was leaking from the lapjoint area when the catheter was flushed.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Event Description
Reportable based on device analysis completed on 14mar2016.It was reported that catheter kink occurred.The target lesion was located in the severely tortuous and mildly calcified left main coronary artery.An opticross¿ imaging catheter was used to view the target lesion.During percutaneous coronary intervention (pci), the opticross¿ imaging catheter was advanced; however, it was noted that the exit port part of the device got kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's condition is good.However, device analysis revealed an open hole in the lapjoint area.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS¿
Type of Device
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS  
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key5548537
MDR Text Key41939980
Report Number2134265-2016-02513
Device Sequence Number1
Product Code OBJ
Combination Product (y/n)N
PMA/PMN Number
K123621
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 03/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/14/2016
Device Model NumberH749518080
Device Catalogue Number51808
Device Lot Number18493619
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/14/2016
Initial Date FDA Received04/05/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/16/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-