• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AESPIRE 7100; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AESPIRE 7100; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 01/27/2016
Event Type  malfunction  
Manufacturer Narrative
No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.A ge healthcare service representative performed a checkout of the equipment and confirmed the reported complaint.The n2o flowmeter needle valve kit, o-rings and spacer parts were replaced.The secondary regulators were calibrated, and the unit was returned to service.
 
Event Description
The hospital reported, during pre-use checkout, that the unit was able to deliver n2o without o2, resulting in a hypoxic mixture.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESPIRE 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda drive
madison WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda drive
madison WI 53718
Manufacturer Contact
stephanie cass
3000 n. grandview boulevard
waukesha, WI 53188-1696
MDR Report Key5550144
MDR Text Key41961681
Report Number2112667-2016-00660
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K000706
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 04/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/08/2016
Initial Date FDA Received04/06/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-