MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG HLM TUBING SET W/SOFTLINE COATING; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG HLM TUBING SET W/SOFTLINE COATING; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HQV 59600
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/08/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A supplemental medwatch will be submitted when additional information becomes available.The component is a vhk 71000 venous hardshell cardiotomy reservoir; 510(k) #k090534.
 
Event Description
It was reported that during use on a patient, when beginning the aortic aspiration, a jet of blood was observed.The luer between the tubing and the cardiotomy reservoir was broken.The affected luer was cut and the tubing connected to another outlet of the reservoir.No clinical consequence was reported.(b)(4).
 
Manufacturer Narrative
Sample is not available for investigation, only a picture was sent with broken luer between tube and reservoir.Thus the failure could be confirmed.Most possible root cause could not be determined as the failure is first of its kind and cannot be investigated.The supplier investigation shows that device history record of the complaint lot has been investigated and no abnormality was found.No further information could be provided by the (b)(4).The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
 
Event Description
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLM TUBING SET W/SOFTLINE COATING
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key5551483
MDR Text Key42623336
Report Number8010762-2016-00246
Device Sequence Number1
Product Code DWF
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K090534
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/01/2017
Device Model NumberHQV 59600
Device Catalogue Number701031267
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/29/2016
Initial Date FDA Received04/06/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/02/2017
Date Device Manufactured03/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-