Model Number HQV 59600 |
Device Problems
Break (1069); Fluid/Blood Leak (1250)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/08/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).A supplemental medwatch will be submitted when additional information becomes available.The component is a vhk 71000 venous hardshell cardiotomy reservoir; 510(k) #k090534.
|
|
Event Description
|
It was reported that during use on a patient, when beginning the aortic aspiration, a jet of blood was observed.The luer between the tubing and the cardiotomy reservoir was broken.The affected luer was cut and the tubing connected to another outlet of the reservoir.No clinical consequence was reported.(b)(4).
|
|
Manufacturer Narrative
|
Sample is not available for investigation, only a picture was sent with broken luer between tube and reservoir.Thus the failure could be confirmed.Most possible root cause could not be determined as the failure is first of its kind and cannot be investigated.The supplier investigation shows that device history record of the complaint lot has been investigated and no abnormality was found.No further information could be provided by the (b)(4).The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
|
|
Event Description
|
(b)(4).
|
|