The customer reported to a ge healthcare (gehc) technical engineer that an exam performed on (b)(6) 2016 exceeded the dose threshold limit of 5gy.No injury has been reported.Initial analysis indicated that the patient received a peak skin dose of 7.57 gy in the same body area during the exam procedure.Further gehc engineering investigation analysis was performed using information from the gehc field engineer and logs from the system which showed that the total exam duration for this exam was 2 h 3 min with a total dose of 7.571 gy, 6.046 gy from fluoroscopy acquisitions and 1.525 gy from digital subtraction angiography (dsa) acquisitions, and the peak incident dose was 4.6 gy.The most probable root cause for the high dose would be a long exam performed on a large patient with appropriate acquisition settings.Acquisition settings were selected by the user based on the physician's clinical judgment for obtaining the best image quality for diagnosis.The patient did not sustain any radiation induced injury.Because the localized skin peak dose was less than 7.5 gy, the likelihood of permanent radiation damage is remote.This system is not covered under a gehc service contract.All system calibrations were completed successfully and there was no indication of any system malfunction.Detailed instructions on reducing dose and improving image quality are provided in the innova igs systems cardiovascular imaging system operator manual.These instructions have been reviewed and found to be appropriate.No system malfunction has been identified.The system worked as intended and per specifications.No further action is required.
|