• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH IMPACTION HANDLE; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH IMPACTION HANDLE; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 6541-4-810
Device Problems Degraded (1153); Material Discolored (1170); Material Integrity Problem (2978)
Patient Problem No Patient Involvement (2645)
Event Date 03/16/2016
Event Type  malfunction  
Manufacturer Narrative
When completed, the investigation results will be submitted in a supplemental report.
 
Event Description
Product handles were reported to be discoloured and degrading.
 
Manufacturer Narrative
A review of the device history records indicates that the reported devices were manufactured and accepted into final stock with no reported discrepancies.The complaint history review indicated that there were no similar events for the reported lot.Visual inspection: the device was returned in used condition.Visual inspection confirmed that the green santoprene handle was degraded.Conclusion: the event was confirmed; the device handle is discolored and damaged.The sales representative indicated that the hospital has been experiencing some sterilization issues and using a metal cleaning chemical which could have been a contributing factor to the damage found.The damaged device was discovered during inspection; there was no surgical procedure associated with the reported event.This event meets the definition of preventive maintenance; no further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
Product handles were reported to be discoloured and degrading.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPACTION HANDLE
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
cindy chuhinko
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key5563635
MDR Text Key42198887
Report Number0002249697-2016-01158
Device Sequence Number1
Product Code JWH
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K143393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6541-4-810
Device Lot NumberCR1S01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/16/2016
Initial Date FDA Received04/08/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/15/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-