• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC HEARTWARE® VENTRICULAR ASSIST SYSTEM; CIRCULATORY ASSIST SYSTEM, BATTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC HEARTWARE® VENTRICULAR ASSIST SYSTEM; CIRCULATORY ASSIST SYSTEM, BATTERY Back to Search Results
Catalog Number 1407DE
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/18/2016
Event Type  malfunction  
Manufacturer Narrative
The instructions for use (ifu) and patient manual include a reference guide for both visual and tone alarms including potential causes and actions to take.Additionally there is a warning to keep spare, fully charged batteries and back up controller available at all time.The steps for exchange of batteries and controllers are outlined.The instructions for use and patient manual include a warning to keep a spare back up controller available at all times and outlines that if there is a controller failure, the controller should be switched to the back-up controller.The steps for exchange of devices are also outlined.It further warns that damaged equipment should be reported to the manufacturer and inspected.This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Heartware will submit a supplemental report when new facts arises which materially alters information submitted in a previous mdr report.
 
Event Description
It was reported that during the smr update, it was recognized that display is not working properly and not displaying all dots on the display.Controller was replaced and no consequences to the patient.No additional information provided.
 
Manufacturer Narrative
It was reported that the controller's display was not working properly.The controller, (b)(4), was returned and sent to the supplier for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.The reported event could not be confirmed.Analysis of the device revealed that the device met specifications; the device passed visual examination and functional testing.The results revealed that the controller's liquid crystal display (lcd) was able to display all characters properly.No failure detected; the controller was able to display all characters properly.Heartware will submit a supplemental report when new facts arises which materially alters information submitted in a previous mdr report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE® VENTRICULAR ASSIST SYSTEM
Type of Device
CIRCULATORY ASSIST SYSTEM, BATTERY
Manufacturer (Section D)
HEARTWARE, INC
14400 nw 60th avenue
miami lakes FL 33014 3105
Manufacturer (Section G)
HEARTWARE, INC
14400 nw 60th avenue
miami lakes FL 33014 3105
Manufacturer Contact
nathalie nunez
14400 nw 60th avenue
miami lakes, FL 33014-3105
3053641538
MDR Report Key5564105
MDR Text Key42215490
Report Number3007042319-2016-01552
Device Sequence Number1
Product Code DSQ
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/30/2014
Device Catalogue Number1407DE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/18/2016
Initial Date FDA Received04/08/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/24/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age45 YR
Patient Weight96
-
-