Device identification: the reported device was confirmed to be an probe- sharp straight, p/n 110810, lot 06010113.Device history review: review of the device history records indicate (b)(4) devices were inspected and accepted into final stock on 2/27/2013, with no related nonconformances.An additional part was inspected and released on 7/22/013.Device evaluation and results: visual inspection.Visual inspection shows that the tip of the probe is bent.Dimensional inspection: dimensional inspection was not completed because the visual inspection confirmed the failure.Functional inspection: a robot was used to test the function of the probe.A probe check was completed 5 times, with the average error 1.1 mm.This is just outside an "acceptable" result (> 1 mm) and is considered "not optimal." complaint history review: a review of complaints related to p/n 110810, lot number 06010113 shows 1 additional complaints related to the failure in this investigation.It is (b)(4) and is on the same pi as the current complaint.Tracking of complaints related to the 110810 part number will be tracked through quarterly trend request #860.Conclusions: the failure mode was confirmed.The tip is bent and the probe check produced a "not optimal" result.Corrective action/preventive action: no action is required at this time as there is no indication to suggest a product non-conformity or unanticipated hazard.
|