Catalog Number 00249001200 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Information was received from a distributor who is not required to complete form 3500a.Device return condition noted impact damage on the impaction zones on top and bottom of the product.Threads were damaged and a few scratches were along the side of the product.The device met print specifications where measured.For functional testing, 3.0 and 2.4 mm guide wires were placed in a vice and the guide wire gripper was locked down on the guide wire and hit with a mallet.In both cases the guide wire gripper slid down the guide wire.Device history records review found lot conformed at the time of manufacture.The approximate lifetime of the device is 4 years.It is unknown how many times device was used during that timespan.The device is used for treatment.The root cause of the complaint appears due to normal wear to the device over its approximate 4 year lifetime.The device reached the end of its useful life.This mdr was identified during an internal retrospective review to meet reporting requirements and therefore is being filed retrospectively.
|
|
Event Description
|
It was reported during surgery, the guide wire gripper was not functioning properly and would not tension wires as expected.Cannot be properly jammed and slides through guide wires.In addition, subcomponents on the instrument appear to be loosened.
|
|
Search Alerts/Recalls
|