(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.The assignable root was determined to be due to improper handling and wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
It was reported from (b)(6) that during service and evaluation, it was observed that the battery handpiece device was worn due to insufficient sealing, had a broken trigger/slider, was torn off over running in reverse mode and had the magnetic sensor of the upper trigger broken.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|