As reported, patient has a ppn - put catheter in approx.2 weeks ago.Patient was saying it "wasn't working." patient allergic to contrast, so guidewire was inserted to check catheter.When the wire came out there was a "tear in the wire 2cm in length." picc was removed and replaced.No patient injury was reported.
|
The reported complaint description cannot be confirmed as no product was returned for evaluation.Additionally, a device history records review was unable to be conducted as no lot number was provided.The angiodynamics complaint report was reviewed for the bioflo picc product family and the failure mode, "guidewire unravelled." no adverse trend was identified.Angiodynamics' process controls related to the bioflo picc are as follows: incoming inspection of tapered extruded tubing: a visual inspection based on an aql, which includes but is not limited to, visualizing for dents in the catheter tubing is performed.Dents you can feel, but cannot see are acceptable; dents you can feel and see are rejectable.In addition, a guidewire insertion test is performed.The guidewire must pass through the entire length of the extrusion with no resistance.Additionally, manufacturing performs an in-process visual inspection for molding defects and a guidewire insertion test to verify lumen patency, at defined intervals throughout the molding process of each work order.The guidewire must pass through the catheter with little to no resistance to be considered acceptable.At this time, angiodynamics has deemed their process controls adequate to address the reported failure mode.(b)(4).Device not returned to manufacturer.
|