Medtronic received information that the distal end of the pipeline accidentally dislodged into the aneurysm.The patient was treated for an unruptured, fusiform, wide-necked (14 mm) aneurysm located in the left vertebral artery aneurysm arising proximal of the pica.It was reported that the pipeline was initially deployed nicely across the aneurysm.Upon removal of the pushwire and advancement of the microcatheter to capture distal protective coil, the device was accidentally dislodged into the aneurysm.The physician decided to telescope another pipeline into the first device to cover the distal segment.The patient is doing fine as timely preventative action taken by physician.Post angiographic results showed slow flow within aneurysm with good stasis.No patient injury was reported as a result of this procedure.
|