Model Number SC-8216-70 |
Device Problem
Material Separation (1562)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 12/11/2013 |
Event Type
Injury
|
Manufacturer Narrative
|
Udi= (b)(4).Device evaluation indicated that the distal electrodes #3r, 7r, 8r were partially dislodged from the paddle silicone, but still attached to their respective cables.Electrode #6r was not returned.Device evaluation indicated that the lead was cut approximately 3.5 inches from the distal end and could not be tested.The damage to the lead is consistent with damages done during the explant procedure and are not considered a failure.
|
|
Event Description
|
A report was received that during device analysis, it was revealed that some electrodes were partially dislodged and one electrode was not returned.
|
|
Manufacturer Narrative
|
Additional information was received that everything was removed.Nothing was left inside the patient's body.
|
|
Event Description
|
A report was received that during device analysis, it was revealed that some electrodes were partially dislodged and one electrode was not returned.
|
|
Search Alerts/Recalls
|