• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER, TURPEAUX INDUSTRIAL PARK MULTIPOLAR BIPOLAR CUP SHELL; KWY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER, TURPEAUX INDUSTRIAL PARK MULTIPOLAR BIPOLAR CUP SHELL; KWY Back to Search Results
Catalog Number 00500104700
Device Problems Nonstandard Device (1420); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/28/2014
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The bipolar metal shell was not returned as it was implanted inside the patient; however, the low density polyethylene (ldpe) bag used for packaging of the device was returned for review.Photo and visual exams revealed the polyethylene bag returned has a hole and indicates package adhesion to the product it contained.The device is used for treatment.This issue has been previously investigated and as part of corrective action, a new supplier for the ldpe bags has been selected and testing completed to ensure thermal reliability and stability of the new bags.Field action z-0845-2016 was initiated on january 11, 2016 to remove remaining affected units from the field.This field action contains the related part and lot number.This mdr was identified during an internal retrospective review to meet reporting requirements and therefore is being filed retrospectively.
 
Event Description
During surgery, it was noted an implant was found with polyethylene material adhered onto cup surface of the acetabular component.The surgeon completed the surgery with the implant as the material was successfully removed from the component.Only the polyethylene bag was provided for further review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIPOLAR BIPOLAR CUP SHELL
Type of Device
KWY
Manufacturer (Section D)
ZIMMER, TURPEAUX INDUSTRIAL PARK
rt. #1, km 123.4
bldg. #1
mercedita PR 00715
Manufacturer (Section G)
TURPEAUX INDUSTRIAL PARK
rt. #1, km 123.4
bldg. #1
mercedita PR 00715
Manufacturer Contact
kevin escapule
p.o. box 708
warsaw, IN 46581-0708
8006136131
MDR Report Key5573041
MDR Text Key43057580
Report Number2648920-2016-00155
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 04/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Catalogue Number00500104700
Device Lot Number62018447
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/28/2014
Initial Date FDA Received04/13/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/23/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0845-2016
Patient Sequence Number1
-
-